The Awareness Technology 2600 Microplate Re performs fully automatic reading-
The 2600 Microstrip Reader uses biochromatic optics. The standard model includes 4 filters with the option of 6 filters at a nominal charge.
On board data reduction calculations include regressions- curve fitting and cutoff absorbance calculations with step-by-step user prompting and the large non volatile memory can store up to 36 user tests.
Communication with an external device can be completed using the included serial and parallel ports.
| System | |
| Display | 2 Line LCD Alpha Numeric 24 Character LCD |
| Keyboard | 16-key- domed membrane switch- 4 x 4- enunciating |
| Power Requirements | 115V or 230V AC- 50-60Hz (switch selectable) |
| Non-volatile Memory | Battery supported RAM- holds approx. 50 wash protocols |
| System | |
| Residual Volume: | less than 3ul |
| Dispense Accuracy | less than 3 percent CV at 300ul |
| Processing Time (Full Plate) | 55 seconds aspirate/dispense |
| Liquid Contact Materials | Stainless steel- Delrin- silicone- nylon- PVC- polypropylene |
| 8 Probe Wash Head Standard (12 or 16 Probe Wash Heads Available) | |
| 6-VIS: 405- 450- 492- 545- 600- and 630nm | |
| 6-UV: 340- 405- 450- 492- 545- and 630nm | |
| Filter Types | Multiple cavity hard coat interference. 10nm half bandpass |
| Vessel | Standard 96-well microtiter plates- or strip trays |
| Dimensions | |
| L x W x D inch | 15.5 x 13.5 x 7.5 |
| Weight | 22 Ilbs |