You are here: New and Used Lab Equipment Analytical Biotech Flow Cytometers Beckman Epics XL Flow Cytometer

New and Used Lab Equipment Canada, Toronto, Vancouver, Montreal, Calgary - GMI Inc.

Biotech arrow Flow Cytometers arrow Beckman Epics XL Flow Cytometer

Beckman Epics XL Flow Cytometer

  • Analyze up to 4 colors of immunofluorescence from a single air cooled laser
  • Applications include DNA analysis/ platelet studies/ reticulocyte enumeration and cell biology/functional studies
  • Flexibile enough to change filter elements for versatility in research settings
  • State-of-the-art Digital Signal Processing (DSP) for reliable linearity and drift-free amplification and compensation.
  • The single laser design eliminates concerns regarding multi-beam stability- signal delay and alignment.
... Full Description >>>


Complete Product Description

The Beckman Coulter Epics XL Flow Cytometer system features the capability to analyze up to 4 colors of immunofluorescence from a single air cooled laser. Other multi-color applications include: multiparametric DNA analysis- platelet studies- reticulocyte enumeration- cell biology/functional studies and as well as a broad range of research applications.

The EPICS XL flow cytometer combines the analytical power of a research cytometer into a compact- truly walk away- clinical analyzer ergonomically designed to expeditesample throughput. GMI brings you over 20 years combined experience in providing this cytometer at a compelling pricepoint.

The instrument is self contained and biohazard safe. The user retains the flexibility to change filter elements for versatility in research settings. For high volume users- the XL-MCL system offers walk-away sample handling with the Multi Carousel Loader (MCL). Capable of throughputs of up to 100 samples per hour- the fully featured XL-MCL incorporates positive bar-code identification and true vortex mixing prior to aspiration.

The Epics XL remains the only flow cytometer to offer state-of-the-art Digital Signal Processing (DSP) for reliable linearity and drift-free amplification and compensation. The single laser design eliminates concerns regarding multi-beam stability- signal delay and alignment. Compensation is achieved by a unique digital compensation matrix.

The EXPO32 ADC software provides both acquisition and analysis options. This comprehensive software package gives you the power to simplify reporting- speed up data acquisition and analysis- track instrument use and get the most out of your data.

The COULTER EPICS XL-MCL is a laser-based flow cytometer which utilizes color-differentiated fluorescence and light scatter measurements to analyze cells. Cells in liquid suspension are presented under pressure to a flow cell where they are surrounded by a laminar sheath of particle-free liquid. This coaxial stream passes through a flow chamber as a jet. Cells are presented one at a time to a laser beam generated by a 488 nm argon laser. The instrument can handle up to 10-000 cells per second. As the cells pass through the laser beam they will (1) scatter light and/or (2) fluoresce- due to either autofluorescence or to the presence of fluorescent stains which have been selectively absorbed or bound by the cell.

The scatter and fluorescence emissions are collected by detectors (photo diodes and photomultiplier tubes) which convert the signals to voltage pulses that are proportional to the amount of light scattered and/or to the intensity of the fluorescence. These pulses are then amplified and converted to a digital form which can be displayed in numeric and/or histogram formats. This instrument has four PMT?s allowing an investigator to look at four colors of fluorescence at once.

For high volume users- the XL-MCL system offers walk away sample handling for superior efficiency with the Multi Carousel Loader (MCL). Capable of throughputs of up to 100 samples per hour- the fully featured XL-MCL incorporates positive bar-code identification and true vortex mixing prior to aspiration.

The XL remains the only flow cytometer to offer state-of-the-art Digital Signal Processing (DSP) for reliable linearity and drift-free amplification and compensation. The single laser design eliminates concerns regarding multi-beam stability- signal delay and alignment. Compensation is achieved by a unique digital compensation matrix.

Beckman Coulter Epics XL Flow Cytometer System Specifications

General  
Operation Color Option 3-color fluorescence with 4-color option
Cooling Air-cooled 488nm 15mw laser
Beam 10u x 90u
Detector Range 300nm-800nm
Flow-Cell Size 250u x 250u BioSense
Analysis Range 3300 cells/sec minimum
Min Cell Size- um 0.5 Maximum
Max Analysis Rate Greater than 3300 cells/sec
Color Dot Density Contour/ isometric/ tomogram/ overlay/ legend and projection plots/ PRISM displays- color precedence or color-blend gating- automatic Boolean logic gate creation- 32 gates per file- real-time statistics.
Dimensions  
XL (W x D x H) inch's 24 x 22.5 x 20
XL-MCL (W x D x H) inch's 34.1 x 22.5 x 20
XL Weight Il bs 140 Il bs
XL-MCL Weight Il bs 187 Il bs