The ChemWell 2910 Immunology and Clinical Chemistry Analyzer is a fully automated analyzer capable of running both EIAs and general chemistries in standard microwells.
ChemWell is a completely open system that is easy to program. Set up assays, routine jobs, quality control, panels, and even index calculations to suit your laboratory. With a completely open system you can program additional wash volume, increase the number of washes, or even direct the probe to pick up and dispense a probe cleaning solution after each specimen. And it is entirely password protected.
Precise pipetting of low volume specimens, elimination of carryover, and careful control of temperature are all essential for reliable results. ChemWell 2910 Immunology and Clinical Chemistry Analyzer can make pre-dilutions, however, this is often not necessary since a high level of precision is achieved even with a 2uL sample. ChemWell 2910 Immunology and Clinical Chemistry Analyzer can process EIA at ambient temperature or 37C. An incubator is provided for biochemistry wells to provide precise kinetic reactions. The probe is also temperature controlled to deliver 37C reagents.
RCA - Reagent Cooling Accessory now available for ChemWell. Protects reagents from ambient heat. The removable racks can be refrigerated, pre-loaded, and ready for the next use.
ChemWell is designed to maximize dependability, versatility, and economy.
ChemWell uses standard, off-the-shelf microwells for all reactions. Performing biochemistry assays in microwells offers many advantages such as increased throughput and decreased cost per test. Uncoated microwells make low volume sample cups for biochemistry. They can also be washed for re-use.
| Overall | |
| Typical throughput | up to 200 endpoint reactions per hour, up to 170 kinetic reactions per hour |
| Dimensions | 36.25" (92.1cm) left to right, 18.75" (47.6 cm) height, 21.5" (54.6 cm) front-to-back Approximate weight 100 lbs (45kg) |
| Reagent and Sampling Dispensing | |
| Capabilities | Dilution, pre-dilution, dispensing single or multiple reagents |
| Probes | 316 stainless steel for maximum reagent compatibility, level sensing |
| Maximum number of specimens | 96 |
| Maximum number of reagents | Typically 27 to 44 or more (you can program reagents to go in the sample rack; assorted replaceable racks and custom designed racks are available for various bottle sizes) |
| Standard reagent rack | 27 |
| Reaction Vessel | Standard micro wells, strips or plates |
| Instrument bottles | 2L wash with low volume warning sensor, 1L rinse (or second wash) with low volume warning sensor, 2L waste bottle with full sensor. 1L priming bottle |
| Incubating, timing and temperature control | |
| EIA Mode | Incubation timing is software controlled; set each row of 8 to time separately or time the whole plate together. |
| Thermal control | Well, probe, and tubing; ambient or 37C (other options also available) |
| Reagant cooling | RCA, Reagent Cooling Accessory optional) cools 12 to 15C below ambient through peltier thermoelectric modules connected to an external controller |
| Washing | |
| Wash head | 8-probe, automatic prime and rinse |
| Programs | Create and run user programmable protocols, aspirate, dispense, soak, mix... |
| Reading | |
| Optical design | Reads absorbance in 4 simultaneous channels, NIST traceable calibration user selected monochromatic or biochromatic results 8 position filter wheels: 340, 405, 450, 505, 545, 600, 630, 700 or custom |
| Interference filters | Long life, hard coat, ion assisted deposition, 10nm typical half bandpass |
| Linear Range | -0.2 to 3.0A |
| Photometer Accuracy | +/- 1% or better |
| Certifications | NRTL Listed, CE Marked |